Setting Measurement Conditions
Sweeping Multiple Sweep Ranges with Different Conditions in a Single
Action (List Sweep)
5. Setting Measurement
Conditions
Value area in the upper left part of the screen.
Step 29. Use one of the following methods with the keys in the ENTRY block to set the pen number
for the segment (an integer between 1 and 6).
Enter the desired value with the numeric keys (
[1] to [6]) and then press [x1].
Turn the rotary knob (
m
mm
m) until the desired value is set.
Press the step keys ([][¯]) to enter the desired value.
There are six pens, from number 1 to number 6, which are set to different colors (the color
mapping can be changed). Accordingly, by assigning a different colored pen to each
segment, you can color-code the traces on the screen to more easily identify each segment.
The factory setting of the graphic drawing pens are: pen 1: white, pen 2: red, pen 3: yellow,
pen 4: green, pen 5: cyan, and pen 6: blue. For how to reset the color of the pens, refer to
“Customizing the Color of Each Screen Item” on page 208.
The pen number that is set for each segment is not displayed on the list sweep table. To
check the pen number, use the
SEGMENT key to specify a segment and press the PEN key.
Step 30. Repeat steps 3 through 27 until you obtain all segments you need. Use the editing functions
below as necessary.
NOTE The number of segments that can be set in a list sweep table ranges from 1 to 18.
The number of points that can be set for a single segment is an integer ranging from 2 to
201. Up to 801 points can be set for each segment.
The sweep frequency range for each segment can be set with a combination of either sweep
start value and sweep stop value or one of sweep center value and sweep span value. Also,
after setting a sweep range with sweep start value and sweep stop value, pressing the keys
for setting sweep center/span value will override the previous format of the sweep range
for the list sweep table. This also works in changing the other direction, from center/span to
start/stop setting values.
The sweep frequency range set for any segment can overlap with any range set for the other
segments.
When appending a segment to the one that is selected, the same values as those in the
Editing function for list sweep table Key stroke
Select a segment number (current selection is
marked with “>” to the left side of the segment
number (SEG row).
SEGMENT - (enter using keys in ENTRY
block)
Enter edit mode for the selected segment EDIT
Delete the selected segment (pressing the key
will delete the segment without confirmation).
DELETE
Append a segment to the selected segment and
enter edit mode.
ADD
Delete all segments of the list sweep table in
edit mode.
CLEAR LIST - yes (cancel to cancel)