Page view 295
288 Chapter 8
Analysis and Processing of Result
Save and Recall the Agilent 4294A Internal Data
The second and later lines are data lines. The contents of the data lines depend on the
setting in the option line.
Saving display screen (GRAPHICS save)
You can save the screen displayed on the 4294A in the TIFF (Tagged Image File Format)
format. It is useful when you want to import measurement result graphs onto software on
your PC and using them.
Step 1. Press the
[Copy] key to display the copy menu.
Step 2. Press the
SETUP key to display the copy setup menu.
Step 3. Select colors.
NOTE Although menu items other than shown in the above table appear, they do not affect image
files you save.
Step 4. Press the
[Save] key to display the save menu.
Step 5. Press the
STORE DEV [ ] key to display the storage device menu.
Step 6. Select one of the storage devices used with the 4294A.
NOTE When you use the floppy disk drive, insert a floppy disk into it.
You cannot save data on a floppy disk that is write-protected. Make the disk write-enabled
before inserting it.
Table 8-21 Contents of data line
Description
First column A frequency. Its unit is kHz specified in the option line.
Second column An absolute value of the S parameter.
Third column A phase of the S parameter.
Colors Key operation
Black and white (default)
B&W
Using default colors FIXED
COLOR
Colors used to display the screen IMAGE
COLOR
Storage device Key operation
Floppy disk drive
FLOPPY
Volatile memory disk MEMORY
Nonvolatile memory disk (flash memory) FLASH MEMORY