Agilent Technologies 4294A Instruction Manual Page 447

  • Download
  • Add to my manuals
  • Print
  • Page
    / 469
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 446
Appendix C 439
Error messages
Memory error
C. Error messages
and a floppy disk is write-protected. If necessary, disable the write-protection of the floppy
disk.
-311 Memory error
An error is detected in the memory of the Agilent 4294A.
34 MEMORY TRACE NOT ACTIVE
Although no data is stored in the memory trace, you attempt to execute a command (DISP
MEMO, DMNM, and so on) that uses the memory trace. Therefore, the command is
invalid. First, use the DATMEM command and store data in the memory trace.
-252 Missing media
You attempt to access a storage device, but the media type does not exist or is not installed
correctly. This message is displayed, for example, when you specify the floppy disk drive
as the storage device but no floppy disk is correctly inserted into the drive.
-109 Missing parameter
The number of parameters is less than that the command requires. For example, although
the “POIN” command requires 1 parameter such as “POIN 101,” no parameter is added as
POIN.
36 MUST BE MORE THAN 2 POINTS FOR ANALYSIS
Although the calculation of equivalent circuit parameters (CALCULATE PARAMETERS
key, CALC PARMS key, or CALECPARA command) is executed, because the number of
points (NOP) within the sweep range (if the partial search function is on and within the
specified search range) is two, they cannot be calculated. Set the number of points within
the sweep range (if the partial search function is on, within the specified search range) to
three or more.
N
98 NO ACTIVE MARKER
When the marker is not displayed, you attempt to execute the marker moving (Marker?)
function and, as a result, the command is ignored. For example, when the marker is not
displayed, a command corresponding to the MKR?START key (MKRSTAR) is sent. Before
executing the marker moving function, display the marker (command: MKR ON).
0 (No error)
No error has occurred. This message is not displayed on the LCD. 0 is returned as the error
number if no error occurs in the instrument when the OUTPERRO? command is sent
through GP-IB.
100 NO FIXED DELTA MARKER
When the fixed D marker (FIXED DMKR) is not displayed, you attempt to execute a
command of a function that needs the display of the fixed D marker and, as a result, the
command is ignored. For example, when the fixed D marker is not displayed, a command
corresponding to the FIXED DMKR VALUE, FIXED DMKR AUX VALUEN key
(DMKRVAL <value>, DMRRAUV <value>) is sent. Before executing a function using the
fixed D marker, first display the fixed D marker (command: DMKR FIX).
113 NO STATE/DATA FILES ON DISK
Although you attempt to display the list of files and directories on a floppy disk on the
Page view 446