Agilent Technologies 4294A Instruction Manual Page 142

  • Download
  • Add to my manuals
  • Print
  • Page
    / 469
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 141
134 Chapter 6
Calibration
A. Calibration When Using Direct Connection Type Test Fixture
16089B Kelvin Clip Lead
16089C Kelvin Clip Lead
16089D Alligator Clip Lead
16089E Kelvin Clip Lead
Step 1. Make sure that the adapter selection is set to “No Adapter (
NONE).”
See “Adapter Setting” on page 67 for adapter selection.
Step 2. Perform the fixture compensation in accordance with Fixture Compensation” on page
151.
NOTE When you use a direct connection type test fixture, neither user calibration nor port
extension compensation is required for measurement (they cannot be performed).
Page view 141
1 2 ... 137 138 139 140 141 142 143 144 145 146 147 ... 468 469

Comments to this Manuals

No comments