Appendix B 401
Key Definitions
Softkeys displayed by pressing the [Search] key
B. Key Definitions
MKR ® MAXIMUM Substitutes the sweep parameter value at the main marker on the current active
trace (A or B) into the partial search range upper limit. In other words, the sweep
range is changed using the current position of the main marker as the new partial
search range upper limit.
MKRTRMAX
MKRD ® SEARCH
RNG
Uses the range between the main marker on the current active trace (A or B) and
the D marker as the partial search range.
MKRTR
return Returns to Table B-64, “Search (Search) menu,” on page 397. (N/A)
1. When the sweep type is not list (list sweep).
Table B-70 Search range (Search Rng) menu (2)
Key operation:
[Search] - SEARCH RANGE MENU
1
Softkey Function GPIB command
FULL
|
Selects the mode in which the partial search range is the entire list sweep table. SEARNG FULL
SEGMENT Selects the mode in which the partial search range is one specific segment in the
list sweep table.
SEARNG SEGMENT
SEGMENT NUMBER When a segment is selected as the partial search range, specify a segment
number for partial search.
SEGMNUM <value>
RNG LINE on OFF Toggles on and off the partial search range border line display to indicate the
borders of the partial search range by two vertical lines on the graph. If the
measurement parameter is COMPLEX Z-Y, you cannot turn on the partial search
range border line display.
SEARLINE {ON|OFF}
return Returns to Table B-64, “Search (Search) menu,” on page 397. (N/A)
1. When the sweep type is list (list sweep).
Table B-69 Search range (Search Rng) menu (1)
Key operation:
[Search] - SEARCH RANGE MENU
1
Softkey Function GPIB command