Chapter 8 263
Analysis and Processing of Result
Set a limit to the trace and make pass/fail evaluation
8. Analysis and Processing
of Result
Figure 8-27 Limit line test function ON (example of rejection by test)
NOTE When the limit line test is turned on, the test result (pass/fail) is output as a PASS/FAIL
signal through the 24-bit I/O port of the rear panel.
Step 4. Operate the
BEEP FAIL on OFF (or BEEP FAIL ON off) key to activate a beep sound in
the event that the limit line test indicates failure.
NOTE The limit line table can be edited with the limit line test function turned on.
If the limit line test function is turned on, the pass/fail result is indicated by the following
four methods.
“PASS” or “FAIL” is displayed in the Limit Line Test Result field on the right side of
the screen.
If the fail beep setting is On (
BEEP FAIL ON off), a beep sounds in the event of a
failure.
If you execute the list display (
LIST VALUE key) of the measurement parameter value
(measured value) by following the procedures in “Print the measurements and settings
(text)” on page 296, an asterisk “*” indicating a failure is displayed next to
measurement points that exceeds the limit.
Softkey label Beep sound when rejected
BEEP FAIL ON off ON (beeps)
BEEP FAIL on OFF OFF (doesn’t beep)