138 Chapter 6
Calibration
C. Calibration for Four-Terminal Pair, 2-m Extension
Step 2. Perform the fixture compensation in accordance with Fixture Compensation” on page
151.
NOTE When the measurement port is extended by 2 meters with the four-terminal pair
configuration, neither user calibration nor port extension compensation is required for
measurement (they cannot be performed).