Appendix B 379
Key Definitions
Softkeys displayed by pressing the [Cal] key
B. Key Definitions
SETUP ä Displays Table B-36, “Adapter setup (Adapter Setup) menu,” on page 379. If
you select an adapter configuration corresponding to a key other than the NONE
key (HP 16047E test fixture, etc.), you have to enter this menu when you use the
adapter (4TP 1M, 4TP 2M, 7mm 42942A, PROBE 42941A) for the first time
and measure the calibration standard. The setup data of adapters once set up are
stored in the internal non-volatile memory of the 4294A, and therefore re-setup
is not required under normal conditions. If the NONE key is selected as the
adapter configuration, this SETUP key is invalid (setup is not required).
(N/A)
return Returns to Table B-34, “Calibration (Calibration) menu,” on page 378. (N/A)
Table B-36 Adapter setup (Adapter Setup) menu
Key operation: [Cal] - ADAPTER [ ] - SETUP
Softkey Function GPIB command
PHASE COMP [ ] Executes the measurement of the phase compensation data. The content of [ ] is -
before the data acquisition and changes to DONE after the data acquisition. If the
4TP 1M key, 4TP 2M key, 7mm 42942A key, or PROBE 42941A key has been
selected in Table B-35, “Adapter configuration (Adapter) menu,” on page 378,
this phase compensation data must be measured.
ECALP
OPEN [ ] Executes the measurement of the OPEN data. The content of [ ] is - before the
data acquisition and changes to DONE after the data acquisition. If the 7mm
42942A key or PROBE 42941A key has been selected in Table B-35, “Adapter
configuration (Adapter) menu,” on page 378, this OPEN data must be measured.
Otherwise, the measurement of the OPEN data is not required (OPEN [ ] key is
invalid).
ECALA
SHORT [ ] Executes the measurement of the SHORT data. The content of [ ] is - before the
data acquisition and changes to DONE after the data acquisition. If the
7mm 42942A key or PROBE 42941A key has been selected in Table B-35,
“Adapter configuration (Adapter) menu,” on page 378, this SHORT data must
be measured. Otherwise, the measurement of the SHORT data is not required
(SHORT [ ] key is invalid).
ECALB
LOAD [ ] Executes the measurement of the LOAD data. The contents of [ ] is - before the
data acquisition, and changes to DONE after the data acquisition. If the 4TP 1M
key, 4TP 2M key, 7mm 42942A key, or PROBE 42941A key has been selected
in Table B-35, “Adapter configuration (Adapter) menu,” on page 378, this
LOAD data must be measured.
ECALC
done Based on the above acquired data, calculates the setup data for the selected
adapter and stores it in the internal non-volatile memory of the 4294A. For
adapters whose setup data is once acquired, re-setup is not required under normal
conditions. Executing this key returns to Table B-35, “Adapter configuration
(Adapter) menu,” on page 378.
ECALDON
cancel Cancels the setup for the selected adapter and returns to Table B-35, “Adapter
configuration (Adapter) menu,” on page 378. Setup data previously executed is
valid.
ECALQUI
Table B-35 Adapter configuration (Adapter) menu
Key operation: [Cal] - ADAPTER [ ]
Softkey Function GPIB command