Chapter 8 233
Analysis and Processing of Result
Define the Peak
8. Analysis and Processing
of Result
Define peak sharpness
The following steps are used to define the peak sharpness:
Step 1. Press the [Search] key to display the Search menu.
Step 2. Press the
PEAK key to display the Peak menu.
Step 3. Press the
PEAK DEF MENU key to display the Peak Definition menu.
Step 4. Press the
PEAK DEF DX key. The set value of the current peak DX is displayed in the set
parameter value field in the upper left area of the screen.
Step 5. Use the keys or rotary knob of the ENTRY block in one of the following ways to specify a
new peak DX value.
Enter the desired value with the numeric keys (
[0] to [9] and [.]) and then press one of
the unit keys (
[G/n], [M/m], [k/m], or [´1]).
Turn the rotary knob (
m
mm
m) until the desired value is set.
Press the step keys ([][¯]) to set the desired value.
Step 6. Press the
PEAK DEF DY key to display the set value of the current peak DY in the set
parameter value field in the upper left area of the screen.
Step 7. Use the keys or rotary knob of the ENTRY block in one of the following ways to specify a
new peak DY value.
Enter the desired value with the numeric keys (
[0] to [9] and [.]) and then press one of
the unit keys (
[G/n], [M/m], [k/m], or [´1]).
Turn the rotary knob (
m
mm
m) until the desired value is set.
Press the step keys ([][¯]) to set the desired value.
Define peak sharpness using a peak on the trace
A peak on the trace can be used to define peak sharpness.
Step 1. Activate the trace for which the peak sharpness is defined.
Step 2. Move the main marker to the peak used to define the peak sharpness.
NOTE Here, the main marker can be moved to the target peak by a peak search according to
PEAK PLRTY [NEG] Negative (if the measurement at that point is
smaller than the measurement of the adjacent
measuring points, it is regarded as a peak)
Softkey label Selected peak polarity
Active Trace Key operation
Trace A [A]
Trace B [B]