Chapter 5 91
Setting Measurement Conditions
Setting Sweep Range
5. Setting Measurement
Conditions
NOTE After you change the sweep range, the trace for the new sweep range will not be displayed
unless you perform a new sweep measurement.
Figure 5-6 Before
MKR ® CENTER
Set the difference in the sweep parameter value
between the current main marker position and
the delta marker position
1
as the new sweep
span value.
MKR D ® SPAN
Search the trace for a peak
2
and specify a new
sweep center value as the location of the peak
found.
more 1/2 - PEAK ® CENTER
The current main marker position is set as the
new sweep center value, and the new sweep
span value is set depending on the specified
zooming aperture
3
. As a result, the trace will
be magnified (zoomed in) by focusing on the
main marker position.
more 1/2 - ZOOMING APERTURE to set the
zooming aperture, and then
MKR ZOOM
1. For how to use the main marker and the delta marker, refer to Chapter 8 , “Analysis and
Processing of Result,” on page 211.
2. Refer to “Search the maximum/minimum peak” on page 228 and “Define the Peak” on
page 232 for information on the peak search function.
3. For the ratio of the new sweep span value to the current sweep span value (%), you can
set a value between 0.01% and 100% in steps of 0.01%. The power-on and preset set-
tings are both 10%.
Setting sweep range Key stroke