432 Appendix C
Error messages
CALIBRATION NOT USABLE
four-terminal pair extension.
12 CALIBRATION NOT USABLE
A command to execute data measurement for unnecessary (not allowed) calibration (user
calibration or fixture compensation) for the current adapter selection is sent. The command
is invalid. For example, when NONE (no adapter) is selected as the adapter selection,
sending the CALA command, which executes OPEN data measurement for the user
calibration, causes this error (when the adapter selection is NONE, you cannot execute the
user calibration).
Execute the measurement of necessary (allowed) calibration data for the current adapter
selection. For more information on the appropriate adapter selection and calibration, refer
to Table 6-1, “Device Connection and Corresponding Adapter Selection and Calibration,”
on page 130
11 CALIBRATION REQUIRED
In user calibration or fixture compensation, although error coefficients based on the
measurement of the required calibration standards (OPEN, SHORT, or LOAD) have not
been acquired, a GP-IB command (CALST ON, COMCA ON, COMCB ON, or COMCC
ON) to turn on the user calibration function or fixture compensation function is sent. Valid
error coefficients are not detected, and the command is invalid. For user calibration, it is
necessary to measure all calibration standards (OPEN, SHORT, LOAD) and acquire error
coefficients based on the measurement. On the other hand, for fixture compensation, it is
necessary to measure a calibration standard (OPEN, SHORT, or LOAD) corresponding to
the fixture compensation function (OPEN compensation function, SHORT compensation
function, or LOAD compensation function) you select and obtain error coefficients based
on the measurement.
-281 Cannot create program
Programs cannot be created. Insufficient memory is suspected.
35 CAN'T CALCULATE EQUIVALENT PARAMETERS
Because measurement data does not match the selected equivalent circuit, equivalent
circuit parameters cannot be calculated. Acquire data again or select another equivalent
circuit.
82 CAN'T CHANGE—ANOTHER CONTROLLER ON BUS
When an active controller is on the same GP-IB bus, you attempt to set the Agilent 4294A
as a system controller. Unless another active controller is removed from the same GP-IB
bus, you cannot set the 4294A as a system controller.
133 CAN'T CHANGE ON LIST SWEEP
When the sweep type is set to the list sweep, you attempt to execute a command to specify
the lower limit of a partial search range (SEARMIN) or a command to specify the upper
limit of a partial search range (SEARMAX) and, as a result, the command is ignored.
When the sweep type is the list sweep, you cannot specify a partial search range by using
the lower limit and the upper limit. You must first specify a sweep type other than the list
sweep (command: SWPT LIN or SWPT LOG).
130 CAN'T CHANGE ON ZERO SPAN
During the execution of the list sweep when the sweep span of each segment is zero, you
attempt to change the list sweep span (integrated total span of individual segments
specified as the horizontal axis of the graph for the list sweep) to a single span (specifying
a linear frequency axis for the horizontal axis of a graph whose left edge and right edge are