Agilent Technologies 4294A Instruction Manual Page 247

  • Download
  • Add to my manuals
  • Print
  • Page
    / 469
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 246
Chapter 8 239
Analysis and Processing of Result
Analyze trace bandwidth
8. Analysis and Processing
of Result
Step 4. Select the definition of the cutoff point in trace bandwidth analysis from the following:
Step 5. When the fixed value is selected (using the
FIXED VALUE key) in step 4, the fixed value
(measurement parameter value) to define the current cutoff point is displayed in the set
parameter value field in the upper left area of the screen.
Step 6. Use the keys or rotary knob of the ENTRY block in one of the following ways to specify
the fixed value of the cutoff point.
Enter the desired value with the numeric keys (
[0] to [9], [-]and [.]) and then press one
of the unit keys (
[G/n], [M/m], [k/m], or [´1]).
Turn the rotary knob (
m
mm
m) until the desired value is set.
Press the step keys ([][¯]) to set the desired value.
To return to the search menu after defining the cutoff point, press the
return key.
Setting the delta marker in the trace bandwidth analysis
The definition of each parameter obtained in the trace bandwidth analysis varies depending
on the delta marker mode selected at the time. Before executing the trace bandwidth
analysis, select the delta marker mode. If you use the fixed delta marker mode, set the delta
marker position also.
Step 1. Activate the trace for which you want to perform the trace bandwidth analysis.
Step 2. Press the
[Marker] key to display the Marker menu.
Step 3. Press the
DMODE MENU key to display the Delta Marker menu.
Step 4. Select the delta marker mode that specifies the definition of each parameter in the trace
Definition of cutoff point
1
1. MKRVAL: Indicates the measurement parameter value (measurement) at the position of
the main marker (marker 0).
Fixed value: Indicates the absolute value of the measurement parameter if there is no D
marker; it is a relative value with reference to the tracking D marker if tracking (D
marker that moves with the main marker) is present and a relative value with reference
to the fixed D marker if the fixed D marker is present.
Key operation
MKRVAL/(Ö2)
MKRVAL*(Ö2)
MKRVAL/2
Fixed value (specified by numerical value) FIXED VALUE
MKRVAL
2
-------------------------
MKRVAL 2´
MKRVAL
2
-------------------------
Active Trace Key operation
Trace A
[A]
Trace B [B]
Page view 246
1 2 ... 242 243 244 245 246 247 248 249 250 251 252 ... 468 469