Chapter 8 251
Analysis and Processing of Result
Calculate the equivalent circuit parameter and simulate the frequency
characteristics
8. Analysis and Processing
of Result
Step 8. Press the CALCULATE PARAMETERS (or CALC PARMS) key to calculate the equivalent
circuit parameter.
When the equivalent circuit parameter is on, the calculated equivalent circuit parameter
will be shown under the equivalent circuit diagram. In the case of measurement trace
display (equivalent circuit diagram is off), the calculated equivalent circuit parameter will
be shown in the equivalent circuit parameter area on the left of the screen.
Figure 8-20 Result of equivalent circuit parameter calculation (using equivalent circuit B)
NOTE When the partial search range is set, the equivalent circuit parameter is calculated based on
the trace within that range. For the partial search range setting procedure, see “Specify
partial search range” on page 267.
The display of the equivalent circuit parameter on the measurement trace screen can be
turned on and off with the
PARAMETERS ON off key (PARAMETERS on OFF). When the
equivalent circuit parameter has been calculated by pressing the
CALCULATE PARAMETER key, the parameter indication is turned on automatically.
E
2
Resonator
1. Measurement parameter: | Z | - q, Sweep type: log, Vertical axis: | Z | is log and q is linear
2. Measurement parameter: | Z | - q, Sweep type: linear (or log), Vertical axis: | Z | is log and q is linear
Table 8-3 Selection of equivalent circuit
Equivalent circuit model Typical frequency characteristics Sample (example)