Agilent Technologies N8201A Service Manual Page 105

  • Download
  • Add to my manuals
  • Print
  • Page
    / 332
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 104
Programming the Status Register System 4
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 105
If the status register is configured to SRQ on end-of-sweep or measurement and the mode
set to continuous, restarting the measurement (INIT command) can cause the measuring
bit to pulse low. This causes an SRQ when you have not actually reached the
“end-of-sweep” or measurement condition. To avoid this, do the following:
1 Send the command INITiate:CONTinuous OFF.
2 Set/enable the status registers.
3 Restart the measurement (send INIT).
Status Register SCPI Commands
Most monitoring of N8211A/N8212A conditions is done at the highest level using the IEEE
488.2 common commands listed below. You can set and query individual status registers
using the commands in the STATus subsystem.
*CLS (clear status) clears the Status Byte Register by emptying the error queue and
clearing all the event registers.
*ESE, *ESE? (event status enable) sets and queries the bits in the Standard Event
Enable Register which is part of the Standard Event Status Group.
*ESR? (event status register) queries and clears the Standard Event Status Register
which is part of the Standard Event Status Group.
*OPC, *OPC? (operation complete) sets bit #0 in the Standard Event Status Register
to 1 when all commands have completed. The query stops any new commands from
being processed until the current processing is complete, then returns a 1.
*PSC, *PSC? (power-on state clear) sets the power-on state so that it clears the
Service Request Enable Register, the Standard Event Status Enable Register, and
device-specific event enable registers at power on. The query returns the flag setting
from the *PSC command.
*SRE, *SRE? (service request enable) sets and queries the value of the Service
Request Enable Register.
*STB? (status byte) queries the value of the Status Byte Register without erasing its
contents.
:STATus:PRESet presets all transition filters, non-IEEE 488.2 enable registers, and
error/event queue enable registers. (Refer to Table 3.)
Table 3 Effects of :STATus:PRESet
Register Value after
:STATus:PRESet
:STATus:OPERation:ENABle 0
:STATus:OPERation:NTRansition 0
:STATus:OPERation:PTRransition 32767