Basic Function Commands 8
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 267
:REFerence:STATe
Supported All
[:SOURce]:POWer:REFerence:STATe ON|OFF|1|0
[:SOURce]:POWer:REFerence:STATe?
This command enables or disables the RF output reference.
ON (1) Sets the power reference state ON.
OFF (0) Sets the power reference state OFF.
Once the reference state is ON, all subsequent output power settings are set relative to the
reference value. Amplitude offsets can be used with the amplitude reference mode.
Example
:POW:REF:STAT 1
The example sets the reference state on.
*RST 0
:STARt
Supported All
[:SOURce]:POWer:STARt <val><unit>
[:SOURce]:POWer:STARt?
Sets the amplitude of the first point in a step or ramp sweep (Option 007).
During an amplitude sweep operation, N8211A/N8212As with Option 1E1 protect the step
attenuator by automatically switching to attenuator hold (ON) mode. The attenuator is
locked at its current setting and the amplitude sweep range is limited to 40 dB.
Example
:POW:STAR -30DBM
The example sets the amplitude of the first point in the sweep to –30 dBm.
*RST Depends on model and option number
Range Refer to “[:LEVel][:IMMediate][:AMPLitude]" on page 269 for the output power
ranges.