Digital Modulation Commands 10
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 321
:IQADjustment:IOFFset
Supported N8212A
[:SOURce]:DM:IQADjustment:IOFFset <val>
[:SOURce]:DM:IQADjustment:IOFFset?
Adjusts the I channel offset value.
The <val> variable is expressed as a percent with 100% equivalent to 500 mV DC at the
input connector. The minimum resolution is 0.025%.
When using this command to minimize the LO feedthrough signal, optimum performance is
achieved when the command is sent after all other I/Q path commands are executed, such
as those that change the internal phase polarity or adjust the modulator attenuator. If other
adjustments are made after minimizing is performed, the LO feedthrough signal may
increase.
This command is effective only if the state of the I/Q adjustment function is set to ON.
Refer to “:IQADjustment[:STATe]" on page 332.
Example
:DM:IQAD:IOFF -30
The example sets the I channel offset to –30%.
*RST +0.00000000E+000
Range 5E1 to +5E1
:IQADjustment:QOFFset
Supported N8212A
[:SOURce]:DM:IQADjustment:QOFFset <val>
[:SOURce]:DM:IQADjustment:QOFFset?
Adjusts the Q channel offset value.
The <val> variable is expressed as a percent with 100% equivalent to 500 mV DC at the
input connector. The minimum resolution is 0.025%.
When using this command to minimize the LO feedthrough signal, optimum performance is
achieved when the command is sent after all other I/Q path commands are executed, such
as those that change the internal phase polarity or adjust the modulator attenuator. If other
adjustments are made after minimizing is performed, the LO feedthrough signal may
increase.