Agilent Technologies N8201A Service Manual Page 287

  • Download
  • Add to my manuals
  • Print
  • Page
    / 332
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 286
Analog Commands 9
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 287
Sets the internal FM rate using the <val><units> variable, or changes the FM rate by a
user-defined up/down step value. Refer to the
“:FM:INTernal:FREQuency:STEP[:INCRement]" on page 283 for setting the value associated
with the UP and DOWN choices.
The command changes:
the FM rate of the first tone of a dual–sine waveform
the starting FM rate for a swept–sine waveform
the FM rate for all other waveforms
Refer to “:FM[1]|2:INTernal[1]|2:FUNCtion:SHAPe" on page 288 for the waveform
selection.
Example
:FM2:INT:FREQ 40KHZ
The example sets the modulation rate for FM path 2 to 40 kHz.
*RST +4.00000000E+002
Range Dual-Sine & Sine: 0.5 Hz - 1 MHz Swept-Sine 1 Hz to 1 MHz All other
waveforms: 0.5 Hz - 100 kHz
:FM[1]|2:INTernal[1]|2:FUNCtion:NOISe
Supported Option UNT
[:SOURce]:FM[1]|2:INTernal[1]|2:FUNCtion:NOISe GAUSsian|UNIForm
[:SOURce]:FM[1]|2:INTernal[1]|2:FUNCtion:NOISe?
Selects a gaussian or uniform noise type as the modulation. Refer to
“:FM[1]|2:INTernal[1]|2:FUNCtion:SHAPe" on page 288 for the waveform selection.
Example
:FM2:INT2:FUNC:NOIS UNIF
The example selects a uniform noise waveform as the modulation for FM path 2 and FM
source 2.
*RST UNIF
:FM[1]|2:INTernal[1]|2:FUNCtion:RAMP
Supported Option UNT
[:SOURce]:FM[1]|2:INTernal[1]|2:FUNCtion:RAMP POSitive|NEGative
Page view 286
1 2 ... 282 283 284 285 286 287 288 289 290 291 292 ... 331 332

Comments to this Manuals

No comments