System Commands 7
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 207
:QUEStionable:NTRansition
Supported All
:STATus:QUEStionable:NTRansition <val>
:STATus:QUEStionable:NTRansition?
Enables bits in the Data Questionable Negative Transition Filter register. A negative
transition (1 to 0) of corresponding bits in the Data Questionable Condition register will
pass through and be read by the Data Questionable Event register.
The variable <val> is the sum of the decimal values of the bits that you want to enable.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Example
:STAT:QUES:MOD:NTR 3072
This example enables bit 10 (decimal 1024, Baseband is busy) and bit 11 (decimal 2048,
Sweep Calculating) in the Data Questionable Negative Transition Filter register.
Range 0–32767
:QUEStionable:POWer:CONDition
Supported All
:STATus:QUEStionable:POWer:CONDition?
Returns the decimal sum of the bits in the Data Questionable Power Condition register. For
example, if the RF output signal is unleveled (bit 1), a value of 2 is returned.
The data in this register is continuously updated and reflects current conditions.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Range 0–32767
:QUEStionable:POWer:ENABle
Supported All
:STATus:QUEStionable:POWer:ENABle <val>
:STATus:QUEStionable:POWer:ENABle?