28 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
2 Using the LAN Interface
3 Turn on power to the N8211A/N8212A and wait until the LAN LED turns solid green;
this can take up to four minutes depending on whether the instrument is using DHCP or
Auto IP.
Connect to a LAN with a Cross-Over LAN Cable
You can connect the N8211A/N8212A directly to a PC using a crossover cable. To do this,
you should either choose to set IP addresses of the PC and N8211A/N8212A to differ only
in the last digit (example: PC’s IP: 1.1.1.1 and N8211A/N8212A’s IP: 1.1.1.2); or you can use
the DHCP feature or Auto-IP feature if your PC supports them. For more information go to
www.agilent.com, and search on the Connectivity Guide (E2094-90009) or use the Agilent
Connection Expert’s Help to see the Connection Guide.