328 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
10 Digital Modulation Commands
*RST 1
:POLarity[:ALL]
Supported N8212A
[:SOURce]:DM:POLarity[:ALL] NORMal|INVert|INVerted
[:SOURce]:DM:POLarity?
Selects normal or inverted I/Q signal routing. In inverted mode, the Q input is routed to the I
modulator and the I input is routed to the Q modulator, inverting the phase polarity.
NORMal Selects normal routing for the I and Q signals.
INVert (ed) Inverts the phase polarity by routing the I signal to the Q input of the I/Q
modulator and the Q signal to the I input.
Example
:DM:POL INV
The example swaps the I and Q routing paths.
*RST NORM
:SKEW:PATH
Supported N8212A
[:SOURce]:DM:SKEW:PATH RF|BB
[:SOURce]:DM:SKEW:PATH?
Selects the signal path that will be optimized using I/Q skew corrections. The other path
maybe degraded.
RF Optimizes the skew for the I/Q signal applied to the RF Output. The baseband (BB)
output will be functional, but the I/Q skew applied will be optimized for the RF path. While
in real-time mode, the maximum number of user symbols for the FIR is limited to 32.
BB Optimizes the skew for the I/Q signal outputs on the rear-panel. The RF Output will be
functional, but the I/Q skew applied will be optimized for the BB path. While in real-time
mode, the maximum number of user symbols for the FIR is limited to 32.
NOTE
You must have a skew calibration to use this command. I/Q skew corrections and
calibration must be performed at an Agilent factory or service center.