236 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
8 Basic Function Commands
:FREQuency:OFFSet:STATe
Supported All
[:SOURce]:FREQuency:OFFSet:STATe ON|OFF|1|0
[:SOURce]:FREQuency:OFFSet:STATe?
Enables or disables the offset frequency.
Entering OFF (0) will set the frequency offset to 0 Hz.
Example
:FREQ:OFFS:STAT 0
The example disables the frequency offset and sets the offset to 0 hertz.
*RST 0
:FREQuency:REFerence
Supported All
[:SOURce]:FREQuency:REFerence <val><units>
[:SOURce]:FREQuency:REFerence?
Sets the output reference frequency for the N8211A/N8212A. Once the reference
frequency is set, any change to the N8211A/N8212A’s CW frequency will be displayed
referenced to 0 Hz. For example, if the N8211A/N8212A’s CW frequency is set to 100
megahertz and the frequency reference is set (the frequency reference state will
automatically turn on). The frequency display will read 0 Hz. If you change the
N8211A/N8212A’s CW frequency to 1 megahertz, the frequency display will read 1
megahertz. However, the true frequency is 101 MHz. This can be verified by turning the
frequency reference state off. The N8211A/N8212A frequency display will read 101
megahertz. Refer to “:FREQuency:REFerence:STATe for more information.
Example
:FREQ:REF 100MHZ
The example sets the output reference frequency to 100 MHz.
*RST +0.0000000000000E+00
:FREQuency:REFerence:SET