Analog Commands 9
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 283
Frequency Modulation Subsystem ([:SOURce])
:FM[1]|2
Supported All
[:SOURce]:FM[1]|2
Enables the selection of the FM path and is associated with all SCPI commands in this
subsystem.
FM1 FM Path 1 2 with 1 selected
FM2 FM Path 1 2 with 2 selected
When just FM is shown in a command, this means the command applies to path one only.
Each path is set up separately. When a SCPI command uses FM1, only path one is affected.
Consequently, when FM2 is selected, only path two is set up. However, the deviation of the
signals for the two paths can be coupled.
Deviation coupling links the deviation value of FM1 to FM2. Changing the deviation value
for one path changes it for the other. These two paths can be on at the same time provided
the following conditions have been met:
dual–sine or swept–sine is not the selection for the waveform type
each path uses a different source (Internal 1, Internal 2, Ext1, or Ext2)
FM2 must be set to a deviation less than FM1
:FM:INTernal:FREQuency:STEP[:INCRement]
Supported Option UNT
[:SOURce]:FM:INTernal:FREQuency:STEP[:INCRement]<num>|MAXimum|MINimum|
DEFault
[:SOURce]:FM:INTernal:FREQuency:STEP[:INCRement]?
Sets the step value for the internal frequency modulation.
The step value set by this command is used with the UP and DOWN choices for the
command “:FM[1]|2:INTernal[1]|2:FREQuency" on page 286.
The setting enabled by this command is not affected by a power-on, preset, or *RST
command.
Example
:FM:INT:FREQ:STEP 1E5