Agilent Technologies N8201A Service Manual Page 241

  • Download
  • Add to my manuals
  • Print
  • Page
    / 332
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 240
Basic Function Commands 8
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 241
:ROSCillator:BANDwidth:EXTernal
Supported Option UNR/UNX
[:SOURce]:ROSCillator:BANDwidth:EXTernal 25HZ|55HZ|125HZ|300HZ|650HZ
[:SOURce]:ROSCillator:BANDwidth:EXTernal?
Sets the bandwidth of the external reference oscillator.
Example
:ROSC:BAND:EXT 300HZ
The example sets the bandwidth of the external oscillator to 300 Hz.
:ROSCillator:BANDwidth:INTernal
Supported Option UNR/UNX
[:SOURce]:ROSCillator:BANDwidth:INTernal 25HZ|55HZ|125HZ|300HZ|650HZ
[:SOURce]:ROSCillator:BANDwidth:INTernal?
Sets the bandwidth of the internal reference oscillator.
Example
:ROSC:BAND:INT 125HZ
The example sets the bandwidth of the internal oscillator to 125 Hz.
:ROSCillator:SOURce
Supported All
[:SOURce]:ROSCillator:SOURce?
Queries the reference oscillator source: INT (internal) or EXT (external).
:ROSCillator:SOURce:AUTO
Supported Option UNR/UNX
[:SOURce]:ROSCillator:SOURce:AUTO ON|OFF|1|0
[:SOURce]:ROSCillator:SOURce:AUTO?
Enables or disables the ability of the N8211A/N8212A to automatically select between the
internal and an external reference oscillator.
Page view 240
1 2 ... 236 237 238 239 240 241 242 243 244 245 246 ... 331 332