Agilent Technologies N8201A Service Manual Page 288

  • Download
  • Add to my manuals
  • Print
  • Page
    / 332
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 287
288 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
9 Analog Commands
[:SOURce]:FM[1]|2:INTernal[1]|2:FUNCtion:RAMP?
Selects a positive or negative ramp as the internal modulating waveform. Refer to
“:FM[1]|2:INTernal[1]|2:FUNCtion:SHAPe" on page 288 for the waveform selection.
Example
:FM2:INT2:FUNC:RAMP POS
The example selects a positive sloped ramp as the internal modulating waveform.
*RST POS
:FM[1]|2:INTernal[1]|2:FUNCtion:SHAPe
Supported Option UNT
[:SOURce]:FM[1]|2:INTernal[1]|2:FUNCtion:SHAPe SINE|TRIangle|SQUare|
RAMP|NOISe|DUALsine|SWEPtsine
[:SOURce]:FM[1]|2:INTernal[1]|2:FUNCtion:SHAPe?
Selects the FM waveform type. The INTernal2 source selection does not support the
dual–sine or Sweep–Sine waveform types.
Example
:FM2:INT1:FUNC:SHAP SQU
The example selects a square wave as the internal modulating waveform.
*RST SINE
:FM[1]|2:SOURce
Supported Option UNT
[:SOURce]:FM[1]|2:SOURce INT[1]|INT2|EXT1|EXT2
[:SOURce]:FM[1]|2:SOURce?
Selects the FM source.
INT Selects internal source 1 or 2 to provide an ac-coupled signal.
EXT Selects the EXT 1 INPUT or the EXT 2 INPUT connector to provide an externally
applied signal that can be ac- or dc-coupled. The externally applied, ac-coupled input signal
is tested for a voltage level and an annunciator, on the N8211A/N8212A’s front-panel
display, will indicate a high or low condition if that voltage is > ±3% of 1 V