Analog Commands 9
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 295
Example
:LFO:FUNC1:SWE:TRIG EXT
The example sets an external trigger as the trigger for the swept–sine signal.
*RST Free Run
:LFOutput:SOURce
Supported Option UNT
[:SOURce]:LFOutput:SOURce INT[1]|INT2|FUNCtion[1]|FUNCtion2
[:SOURce]:LFOutput:SOURce?
Selects the source for the LF output.
INT Outputs a signal where the frequency and shape of the signal is set by internal
source 1 or 2. For example, if the internal source is currently assigned to an AM path
configuration and AM is turned on, the signal output at the LF OUTPUT connector will have
the frequency and shape of the amplitude modulating signal.
FUNCtion Enables the selection of an internal function generator.
Example
:LFO:SOUR FUNC1
The example selects Function Generator 1 as the active LF source.
*RST INT
:LFOutput:STATe
Supported Option UNT
[:SOURce]:LFOutput:STATe ON|OFF|1|0
[:SOURce]:LFOutput:STATe?
Enables or disables the low frequency output.
Example
:LFO:STAT ON
The example enables the source.
*RST 0