Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 247
:LIST:MODE
Supported All
[:SOURce]:LIST:MODE AUTO|MANual
[:SOURce]:LIST:MODE?
Sets the operating mode for the current list or step sweep.
AUTO Enables the selected sweep type to perform a sweep of all points.
MANual Enables you to select an individual sweep point to control the RF output
parameters. For more about selecting a sweep point, see“:LIST:MANual" on page 246.
Example
:LIST:MODE AUTO
The example sets the mode to automatic.
*RST AUTO
:LIST:POWer
Supported All
[:SOURce]:LIST:POWer <val>{,<val>}
[:SOURce]:LIST:POWer?
Sets the amplitude for the current list sweep points.
The setting enabled by this command is not affected by power-on, preset, or *RST.
During an amplitude sweep operation, N8211A/N8212As with Option 1E1 protect the step
attenuator by automatically switching to attenuator hold mode (OFF). The attenuator locks
at its current setting and the amplitude sweep range is limited to 40 dB. The maximum
number of points is 1601.
Example
:LIST:POW .1,.2,.1,.3,.1,-.1
The example sets the power level for a six point sweep list.
Range See “[:LEVel][:IMMediate][:AMPLitude]" on page 269.