System Commands 7
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 209
Enables bits in the Data Questionable Power Positive Transition Filter register. A positive
transition (0 to 1) of corresponding bits in the Data Questionable Power Condition register
will pass through and be read by the Data Questionable Power Event register.
The variable <val> is the sum of the decimal values of the bits that you want to enable.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Example
:STAT:QUES:POW:PTR 1
This example enables bit 0 (decimal 1, Reverse Power Protection Tripped) in the Data
Questionable Power Positive Transition Filter register.
Range 0–32767
:QUEStionable:POWer[:EVENt]
Supported All
This is a destructive read. The data in the register is latched until it is queried. Once queried,
the data is cleared.
:STATus:QUEStionable:POWer[:EVENt]?
This query returns the decimal sum of the bits in the Data Questionable Power Event
register.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Range 0–32767
:QUEStionable:PTRansition
Supported All
:STATus:QUEStionable:PTRansition <val>
:STATus:QUEStionable:PTRansition?
Enables bits in the Data Questionable Positive Transition Filter register. A positive transition
(0 to 1) of corresponding bits in the Data Questionable Condition register will pass through
and be read by the Data Questionable Event register.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.