Analog Commands 9
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 307
Pulse Modulation Subsystem ([:SOURce])
:PULM:EXTernal:POLarity NORMal:INVerted
Supported Option UNU or UNW
[:SOURce]:PULM:EXTernal:POLarity NORMal|INVerted
[:SOURce]:PULM:EXTernal:POLarity?
Selects the polarity of the TTL input signal at the Pulse In front panel connector. The
N8211A/N8212A can respond to either a normal (a TTL high) or an inverted (TTL low)
signal.
Example
:PULM:EXT:POL NORM
The example selects normal (TTL high) polarity.
*RST Normal
:PULM:INTernal[1]:DELay
Supported Option UNU or UNW
[:SOURce]:PULM:INTernal[1]:DELay <num><time_suffix>|UP|DOWN
[:SOURce]:PULM:INTernal[1]:DELay?
Sets the pulse delay for the internally-generated pulse modulation using the variable
<num>[<time_suffix>]. The command, used with the UP|DOWN parameters, will change
the delay by a user-defined step value. Refer to the “:PULM:INTernal[1]:DELay:STEP" on
page 308 for setting the value associated with the UP and DOWN choices.
The optional variable <time_suffix> accepts nS (nanoseconds) to S (seconds).
The range value is dependent on the pulse period. Refer to “:PULM:INTernal[1]:PERiod" on
page 309 for pulse period settings.
Example
:PULM:INT:DEL 200E-9
The preceding example sets the internal pulse delay to 200 nS.
*RST +0.00000000E+000
Range
Internal Free Run: depends on pulse period and pulse width settings.