Agilent Technologies N8201A Service Manual Page 199

  • Download
  • Add to my manuals
  • Print
  • Page
    / 332
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 198
System Commands 7
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 199
:OPERation[:EVENt]
Supported All
This is a destructive read. The data in the register is latched until it is queried. Once queried,
the data is cleared.
:STATus:OPERation[:EVENt]?
This query returns the decimal sum of the bits in the Standard Operation Event register.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Range 0–32767
:PRESet
Supported All
:STATus:PRESet
Presets all positive and negative transition filters, enable registers, and error/event queue
enable registers.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
:QUEStionable:CALibration:CONDition
Supported All
:STATus:QUEStionable:CALibration:CONDition?
Returns the decimal sum of the bits in the Data Questionable Calibration Condition register.
For example, if the DCFM or DCΦM zero calibration fails (bit 0), a value of 1 is returned.
The data in this register is continuously updated and reflects the current conditions.
Refer to the Chapter 4, “Programming the Status Register System for more information on
programming the status registers.
Range 0–32767
:QUEStionable:CALibration:ENABle
Supported All
:STATus:QUEStionable:CALibration:ENABle <val>
:STATus:QUEStionable:CALibration:ENABle?