330 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz
10 Digital Modulation Commands
:SRATio
Supported N8212A
[:SOURce]:DM:SRATio <val><units>
[:SOURce]:DM:SRATio?
Sets the power level difference (ratio) between the source one and the source two signals
when the two signals are summed together. A positive ratio value reduces the amplitude for
source two while a negative ratio value reduces the amplitude for source one.
The range for the summing ratio is dependent on the modulator attenuator setting for the
N8211A/N8212A that is summing the signals together. The minimum range is achieved
when the modulator attenuator setting is zero and the maximum range is reached when the
maximum attenuator value is used. The range can be calculated using the following
formula:
± Range = 50 dB + Mod Atten
The variable <val> is expressed as a number. The variable <units> is expressed in decibels
(dB).
Example
:DM:SRAT 3DB
The example sets the summing ratio for source 1 and source 2 to 3 dB.
*RST +0.00000000E+000
Range Min: ±50 dB Max: ±90 dB
:STATe
Supported N8212A
[:SOURce]:DM:STATe ON|OFF|1|0
[:SOURce]:DM:STATe?
Enables or disables the internal I/Q modulator. T
The I/Q modulator is enabled whenever a digital format is turned on.
Example
:DM:STAT OFF
The example turns off the I/Q modulator.
*RST 0