7 System Commands
"<file_name,file_type,file_size>"
Refer to “File Name Variables" on page 164 for information on the file name syntax.
:CATalog[:ALL]
Supported All
:MEMory:CATalog[:ALL]?
Outputs a list of all files in the memory subsystem. The return data is in the following form:
<mem_used>,<mem_free>{,"<file_listing>"}
The N8211A/N8212A returns the two memory usage parameters and as many file listings
as there are files in the memory subsystem. Each file listing parameter is in the following
form:
"<file_name,file_type,file_size>"
See Ta ble 1 6 on page 191 for file types, and “File Name Variables" on page 164 for syntax.
:COPY[:NAME]
Supported All
:MEMory:COPY[:NAME] "<src_name>","<dest_name>"
Copies the data from one file into another file. The file can use the same name if the
specified directory is different. For example, if the file resides in non-volatile waveform
memory (NVWFM) it can be copied, using the same name, to the N8211A/N8212A’s
volatile memory (WFM1).
"<src_name>" Names a file residing in memory that will be copied. For information on
the file name syntax, see “File Name Variables" on page 164.
"<dest_name>" TNames the file that is a copy of the "<src_name>" file.
Example
:MEM:COPY "/USER/IQ/4QAM","/USER/IQ/test_QAM"
The example copies the 4QAM file in the N8211A/N8212A’s /USER/IQ directory to a file
named test_QAM and saves it in the same directory.
:DATA:APPend
Supported All
:MEMory:DATA:APPend "<file_name>",<data_block>