Basic Function Commands 8
Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20 / 40 GHz 261
The example starts a single power search of the RF output signal.
*RST 0
:ALC:SEARch:REFerence
Supported All
[:SOURce]:POWer:ALC:SEARch:REFerence FIXed|MODulated
[:SOURce]:POWer:ALC:SEARch:REFerence?
Sets either fixed or modulated modes for power search.
FIXed Uses a 0.5 volt reference.
MODulated Uses the RMS value of the current I/Q modulation as measured during the
power search.
Example
:POW:ALC:SEAR:REF FIX
The example selects a fixed voltage as the reference for a power search.
*RST MOD
:ALC:SEARch:SPAN:START
Supported All
[:SOURce]:POWer:ALC:SEARch:SPAN:STARt <val><units>
[:SOURce]:POWer:ALC:SEARch:SPAN:STARt?
Sets the start frequency for a power search over a user-defined range. The start frequency
has no default value. The start frequency value will be saved before powering off the
instrument.
Example
:POW:ALC:SEAR:SPAN:START 12GHZ
The example selects 12 GHz as the start frequency for a power search.
:ALC:SEARch:SPAN:STOP
Supported All
[:SOURce]:POWer:ALC:SEARch:SPAN:STOP <val><units>
Comments to this Manuals