Agilent Technologies B2902 Service Manual Page 104

  • Download
  • Add to my manuals
  • Print
  • Page
    / 326
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 103
2- 64 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Command Summary
Using Advanced Functions
:STAT:QUES:CAL[:EVEN]?
:STAT:QUES:CURR[:EVEN]?
:STAT:QUES:TEMP[:EVEN]?
:STAT:QUES:TEST[:EVEN]?
:STAT:QUES:VOLT[:EVEN]?
Returns the value of the questionable
calibration, current, temperature, test, or
voltage status event register.
:STAT:QUES:CAL:NTR filter
:STAT:QUES:CURR:NTR filter
:STAT:QUES:TEMP:NTR filter
:STAT:QUES:TEST:NTR filter
:STAT:QUES:VOLT:NTR filter
:STAT:QUES:CAL:NTR?
:STAT:QUES:CURR:NTR?
:STAT:QUES:TEMP:NTR?
:STAT:QUES:TEST:NTR?
:STAT:QUES:VOLT:NTR?
Sets the negative transition filter in the
questionable calibration, current,
temperature, test, or voltage status
register. If you set a bit of the filter, a
1-to-0 transition of its register bit sets the
corresponding bit of the event register.
filter=0 to 65535 (decimal)
0
:STAT:QUES:CAL:PTR filter
:STAT:QUES:CURR:PTR filter
:STAT:QUES:TEMP:PTR filter
:STAT:QUES:TEST:PTR filter
:STAT:QUES:VOLT:PTR filter
:STAT:QUES:CAL:PTR?
:STAT:QUES:CURR:PTR?
:STAT:QUES:TEMP:PTR?
:STAT:QUES:TEST:PTR?
:STAT:QUES:VOLT:PTR?
Sets the positive transition filter in the
questionable calibration, current,
temperature, test, or voltage status
register. If you set a bit of the filter, a
0-to-1 transition of its register bit sets the
corresponding bit of the event register.
filter=0 to 65535 (decimal)
32767