4- 120 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:STATus:<MEASurement|OPERation|QUEStionable>:PTRansition
:STATus:<MEASurement|OPERation|QUEStionable>:
PTRansition
Sets the positive transition filter in the measurement, operation, or questionable
status register. If you set a bit of the filter, a 0-to-1 transition of its register bit sets
the corresponding bit of the event register.
Syntax :STATus:<MEASurement|OPERation|QUEStionable>:PTRansition filter
:STATus:<MEASurement|OPERation|QUEStionable>:PTRansition?
For <MEASurement|OPERation|QUEStionable>, specify MEASurement for the
measurement status register, OPERation for the operation status register, or
QUEStionable for the questionable status register.
Parameter filter Positive transition filter. 0 to 65535 (decimal). Default is 32767.
Parameter data type is NR1 or NDN.
filter is the sum of the binary-weighted values for the set bits.
Query response filter <newline>
filter returns the present setting of the positive transition filter in the specified
register. Response data type is NR1 (decimal) or NDN (binary, octal, or
hexadecimal) selected by the
:FORMat:SREGister command.
Example :STAT:MEAS:PTR 32767
:STAT:QUES:PTR?
:STATus:PRESet
Sets all defined bits in the status system’s PTR registers and clears the all bits in the
NTR and Enable registers. The registers are returned to the default condition.
Syntax :STATus:PRESet
Example :STAT:PRES
:STATus:QUEStionable:<CALibration|CURRent|TEM
Perature|TEST|VOLTage>:CONDition?
Returns the value of the questionable status condition register. See Table 4-7 to 4-11
for the bit definitions. The register setting is not changed by this command.