2- 22 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Command Summary
Setting Source/Measure Unit
Table 2-3 OUTPut Subsystem
Command Summary Reset setting
:OUTP[c]:FILT:AUTO mode
:OUTP[c]:FILT:AUTO?
Enables or disables the automatic filter
function.
mode=1|ON|0|OFF
OFF
:OUTP[c]:FILT[:LPAS]:FREQ freq
:OUTP[c]:FILT[:LPAS]:FREQ? [DEFa
ult|MINimum|MAXimum]
Sets the cutoff frequency of the output
filter. This command setting is ignored if
the automatic filter function is enabled by
the :OUTP:FILT:AUTO command.
freq=MINimum|MAXimum|DEFault|
31.830 Hz to +31.831 kHz
freq = 1/(2 Tconst)
MIN
:OUTP[c]:FILT[:LPAS][:STAT] mode
:OUTP[c]:FILT[:LPAS][:STAT]?
Enables or disables the output filter.
mode=1|ON|0|OFF
ON
:OUTP[c]:FILT[:LPAS]:TCON Tconst
:OUTP[c]:FILT[:LPAS]:TCON? [DEFa
ult|MINimum|MAXimum]
Sets the time constant instead of setting
the cutoff frequency of the output filter.
This command setting is ignored if the
automatic filter function is enabled by the
:OUTP[c]:FILT:AUTO command.
Tconst=MINimum|MAXimum|DEFault|
5 s to 5 ms
Tconst = 1/(2 freq)
MIN
:OUTP[c]:HCAP[:STAT] mode
:OUTP[c]:HCAP[:STAT]?
Enables or disables the high capacitance
mode.
mode=1|ON|0|OFF
OFF
:OUTP[c]:LOW low_state
:OUTP[c]:LOW?
Selects the state of the low terminal.
low_state=FLOat|GROund
GRO
:OUTP[c]:OFF:AUTO mode
:OUTP[c]:OFF:AUTO?
Enables or disables the automatic output
off function.
mode=1|ON|0|OFF
OFF