4- 104 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
[:SOURce]:FUNCtion:TRIGgered:CONTinuous
shape returns DC or PULS. Response data type is CRD.
Example :FUNC PULS
:SOUR2:FUNC:SHAP?
[:SOURce]:FUNCtion:TRIGgered:CONTinuous
Enables or disables continuous trigger output for the specified channel.
Syntax [:SOURce[c]]:FUNCtion:TRIGgered:CONTinuous mode
[:SOURce[c]]:FUNCtion:TRIGgered:CONTinuous?
Parameter mode 0|OFF (default)|1|ON. Parameter data type is boolean.
mode=1 or ON enables continuous trigger output. The specified channel keeps the
output level and range settings even after the grouped channels change status from
busy to idle. The last output settings are saved as the immediate output settings.
mode=0 or OFF disables continuous trigger output. The specified channel changes
the output level and range settings to the previous settings immediately when the
grouped channels change status from busy to idle. The previous settings must be set
by the
[:SOURce]:<CURRent|VOLTage>[:LEVel][:IMMediate][:AMPLitude]
command and the range setup command.
Query response mode <newline>
mode returns 0 or 1, and indicates that continuous trigger is off or on, respectively.
Response data type is NR1.
Example :FUNC:TRIG:CONT 0
:SOUR2:FUNC:TRIG:CONT?
[:SOURce]:LIST:<CURRent|VOLTage>
Sets the source output current or voltage data for the specified channel.
Syntax [:SOURce[c]]:LIST:<CURRent|VOLTage> list
[:SOURce[c]]:LIST:<CURRent|VOLTage>?
For <CURRent|VOLTage>, specify CURRent for current output, or VOLTage for
voltage output.