Agilent B2900 SCPI Command Reference, Edition 2 4- 81
Subsystem Commands
:SENSe:<CURRent[:DC]|RESistance|VOLTage[:DC]>:RANGe[:UPPer]
Syntax :SENSe[c]:RESistance:RANGe:AUTO:ULIMit range
:SENSe[c]:RESistance:RANGe:AUTO:ULIMit? [range]
:SENSe[c]:<CURRent|VOLTage>[:DC]:RANGe:AUTO:ULIMit?
For <CURRent|VOLTage>, specify CURRent for the current measurement, or
VOLTage for the voltage measurement.
Parameter range value (see Table 2-9)|MINimum|MAXimum| DEFault (default
is 200 M). Parameter data type is NRf+. Query does not
support range=value.
Query response range <newline>
range returns the present setting of the maximum measurement range for the auto
range. If a parameter is specified, range returns the value assigned to DEF, MIN, or
MAX. Response data type is NR3.
Example :SENS:RES:RANG:AUTO:ULIM 1E6
:SENS2:CURR:DC:RANG:AUTO:ULIM?
:SENSe:<CURRent[:DC]|RESistance|VOLTage[:DC]>:
RANGe[:UPPer]
Specifies the expected measurement value, and sets the measurement range which
provides the best resolution to measure the specified value.
Syntax :SENSe[c]:<CURRent[:DC]|RESistance|VOLTage[:DC]>:RANGe:UPPer range
:SENSe[c]:<CURRent[:DC]|RESistance|VOLTage[:DC]>:RANGe:UPPer? [range]
For <CURRent[:DC]|RESistance|VOLTage[:DC]>, specify CURRent[:DC] for
current measurement, RESistance for resistance measurement, or VOLTage[:DC]
for voltage measurement.
Parameter range value|UP|DOWN|MINimum|MAXimum|DEFault. Parameter
data type is NRf+. Query does not support range=value, UP,
and DOWN.
value for current measurement: See Table 2-8.
value for voltage measurement: See Table 2-7.