Agilent Technologies B2902 Service Manual Page 244

  • Download
  • Add to my manuals
  • Print
  • Page
    / 326
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 243
4- 122 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:STATus:QUEStionable:<CALibration|CURRent|TEMPerature|TEST|VOLTa
ge>:CONDition?
Table 4-8 Questionable Current Register Bit Definitions
Table 4-9 Questionable Temperature Register Bit Definitions
Table 4-10 Questionable Test Register Bit Definitions
Table 4-11 Questionable Voltage Register Bit Definitions
bit
decimal
value
description definition
0 1 Ch1 Over Current Over current in channel 1.
1 2 Ch2 Over Current Over current in channel 2.
3 to 15 Not used 0 is returned.
bit
decimal
value
description definition
0 1 Ch1 Over Temperature Over temperature in channel 1.
1 2 Ch2 Over Temperature Over temperature in channel 2.
3 to 15 Not used 0 is returned.
bit
decimal
value
description definition
0 1 Ch1 Self-test Failed Self-test failure in channel 1.
1 2 Ch2 Self-test Failed Self-test failure in channel 2.
3 to 15 Not used 0 is returned.
bit
decimal
value
description definition
0 1 Ch1 Over Voltage Over voltage in channel 1.
1 2 Ch2 Over Voltage Over voltage in channel 2.
3 to 15 Not used 0 is returned.
Page view 243