Agilent Technologies B2902 Service Manual Page 191

  • Download
  • Add to my manuals
  • Print
  • Page
    / 326
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 190
Agilent B2900 SCPI Command Reference, Edition 2 4- 69
Subsystem Commands
:READ:ARRay:<CURRent|RESistance|SOURce|STATus|TIME|VOLTage>?
Example :FORM:ELEM:SENS CURR,SOUR
:READ:ARR? (@1,2)
:READ:ARRay:<CURRent|RESistance|SOURce|STATu
s|TIME|VOLTage>?
Executes the :INITiate command and the :FETCh:ARRay:<CURRent|RESistance|
SOURce|STATus|TIME|VOLTage>? command in series, and returns the array data
which contains all of the current measurement data, resistance measurement data,
source output setting data, status data, time data, or voltage measurement data
specified by CURRent, RESistance, SOURce, STATus, TIME, or VOLTage. The
data is not cleared until the :INITiate, :MEASure, or :READ command is executed.
Syntax :READ:ARRay:<CURRent|RESistance|SOURce|STATus|TIME|VOLTage>?
[chanlist]
For <CURRent|RESistance|SOURce|STATus|TIME|VOLTage>, specify CURRent
for current measurement data, RESistance for resistance measurement data,
SOURce for source output setting data, STATus for status data, TIME for time data,
or VOLTage for voltage measurement data.
Parameter chanlist Channels to return the data. Parameter data type is channel list.
(@1)|(@2)|(@1,2)|(@1:2)|(@2,1)|(@2:1). See
“Channel List
Parameter” on page 1-8.
(@1) selects channel 1 only. (@2) selects channel 2 only. (@1,2), (@1:2), (@2,1),
and (@2:1) selects both channels 1 and 2.
If this parameter is not specified, chanlist=(@1) is set.
Query response response <newline>
response returns the array data specified by CURRent, RESistance, SOURce,
STATus, TIME, or VOLTage. Response data type is NR3. See
“Data Output
Format” on page 1-12.
If both channels 1 and 2 are selected by chanlist, response returns the channel 1 data
and the channel 2 data in this order. See the following example. With the ASCII data
output format, each data is separated by a comma.
ch1curr1,ch2curr1,ch1curr2,ch2curr2, ......
ch1curr5,ch2curr5,ch1curr6,+9.910000E+37, ...... ,ch1curr10,+9.910000E+37
Page view 190
1 ... 190 191 192 ... 326

Comments to this Manuals

No comments