4- 158 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:TRIGger<:ACQuire|:TRANsient|[:ALL]>:COUNt
:TRIG2:TRAN:BYP?
:TRIGger<:ACQuire|:TRANsient|[:ALL]>:COUNt
Sets the trigger count for the specified device action.
Syntax :TRIGger[c]<:ACQuire|:TRANsient|[:ALL]>:COUNt trigger_count
:TRIGger[c]<:ACQuire|:TRANsient>:COUNt? [trigger_count]
:TRIGger[c][:ALL]:COUNt? trigger_count
For <:ACQuire|:TRANsient|[:ALL]>, specify :ACQuire for measurement,
:TRANsient for source output, or :ALL for both device actions.
Parameter trigger_count Trigger count. value (1 to 100000)|MINimum|MAXimum|
DEFault (default is 1). Parameter data type is NRf+. Query does
not support trigger_count=value.
Arm count Trigger count must be less than 100001.
Query response response <newline>
response returns the present setting of trigger count. If a parameter is specified,
response returns the value assigned to DEF, MIN, or MAX. Response data type is
NR1.
Example :TRIG:COUN 10
:TRIG2:TRAN:COUN?
:TRIGger<:ACQuire|:TRANsient|[:ALL]>:DELay
Sets the trigger delay for the specified device action.
Syntax :TRIGger[c]<:ACQuire|:TRANsient|[:ALL]>:DELay delay
:TRIGger[c]<:ACQuire|:TRANsient>:DELay? [delay]
:TRIGger[c][:ALL]:DELay? delay
For <:ACQuire|:TRANsient|[:ALL]>, specify :ACQuire for measurement,
:TRANsient for source output, or :ALL for both device actions.