4- 124 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:STATus:QUEStionable:<CALibration|CURRent|TEMPerature|TEST|VOLTa
ge>:NTRansition
For <CALibration|CURRent|TEMPerature|TEST|VOLTage>, specify CALibration
for the questionable calibration status condition register, CURRent for the
questionable current status condition register, TEMPerature for the questionable
temperature status condition register, TEST for the questionable self-test status
condition register, or VOLTage for the questionable voltage status condition register.
Query response value <newline>
value returns the present setting of the specified event register. It is the sum of the
binary-weighted values for the set bits. Response data type is NR1 (decimal) or
NDN (binary, octal, or hexadecimal) selected by the
:FORMat:SREGister
command.
Example :STAT:QUES:CURR:EVEN?
:STAT:QUES:VOLT:EVEN?
:STAT:QUES:TEMP:EVEN?
:STATus:QUEStionable:<CALibration|CURRent|TEM
Perature|TEST|VOLTage>:NTRansition
Sets the negative transition filter in the questionable calibration, current,
temperature, test, or voltage status register. If you set a bit of the filter, a 1-to-0
transition of its register bit sets the corresponding bit of the event register.
Syntax :STATus:QUEStionable:<CALibration|CURRent|TEMPerature|TEST|VOLTage>:N
TRansition filter
:STATus:QUEStionable:<CALibration|CURRent|TEMPerature|TEST|VOLTage>:N
TRansition?
For <CALibration|CURRent|TEMPerature|TEST|VOLTage>, specify CALibration
for the questionable calibration status condition register, CURRent for the
questionable current status condition register, TEMPerature for the questionable
temperature status condition register, TEST for the questionable self-test status
condition register, or VOLTage for the questionable voltage status condition register.
Parameter filter Negative transition filter. 0 to 65535 (decimal). Default is 0.
Parameter data type is NR1 or NDN.
filter is the sum of the binary-weighted values for the set bits.
Query response filter <newline>