2- 32 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Command Summary
Controlling Source/Measure Trigger
Table 2-10 TRIGger Subsystem
Command Summary Reset setting
:ABOR:ACQ [chanlist]
:ABOR:TRAN [chanlist]
:ABOR[:ALL] [chanlist]
Aborts the specified device action for the
specified channel. Trigger status is
changed to idle.
chanlist=(@1)|(@2)|(@1,2)|(@1:2)|
(@2,1)|(@2:1)
(@1) for 1-ch
models
(@1,2) for
2-ch models
:ARM:ACQ[:IMM] [chanlist]
:ARM:TRAN[:IMM] [chanlist]
:ARM[:ALL][:IMM] [chanlist]
Sends an immediate arm trigger for the
specified device action to the specified
channel. When the status of the specified
device action is initiated, the arm trigger
causes a layer change from arm to trigger.
chanlist=(@1)|(@2)|(@1,2)|(@1:2)|
(@2,1)|(@2:1)
(@1) for 1-ch
models
(@1,2) for
2-ch models
:ARM[c]:ACQ[:LAY]:BYP bypass
:ARM[c]:TRAN[:LAY]:BYP bypass
:ARM[c][:ALL][:LAY]:BYP bypass
:ARM[c]:ACQ[:LAY]:BYP?
:ARM[c]:TRAN[:LAY]:BYP?
Enables or disables a bypass for the event
detector in the arm layer.
bypass=ONCE|OFF
ONCE enables the bypass only for the
first passage.
OFF
:ARM[c]:ACQ[:LAY]:COUN count
:ARM[c]:TRAN[:LAY]:COUN count
:ARM[c][:ALL][:LAY]:COUN count
:ARM[c]:ACQ[:LAY]:COUN? [count]
:ARM[c]:TRAN[:LAY]:COUN? [count]
:ARM[c][:ALL][:LAY]:COUN? count
Sets the arm count for the specified
device action.
count=INFinity|MINimum|MAXimum|
DEFault| 1 to 100000 or 2147483647
count=2147483647 indicates infinity.
Query does not support count=INFinity,
1 to 100000 and 2147483647.
Arm count Trigger count must be less
than 100001.