
Agilent B2900 SCPI Command Reference, Edition 2 4- 53
Subsystem Commands
:OUTPut:FILTer:AUTO
OUTPut Subsystem
For the numeric suffix [c], see “Numeric Suffix” on page 1-8.
:OUTPut:FILTer:AUTO
Enables or disables the automatic filter function.
Syntax :OUTPut[c]:FILTer:AUTO mode
:OUTPut[c]:FILTer:AUTO?
Parameter mode 1|ON|0|OFF (default). Parameter data type is boolean.
mode=0 or OFF disables the automatic filter function.
mode=1 or ON enables the automatic filter function. If this function is enabled, the
instrument automatically sets the output filter which provides the best filter
characteristics and cutoff frequency. The following command settings are ignored.
• :OUTPut:FILTer[:LPASs]:FREQuency
• :OUTPut:FILTer[:LPASs]:TCONstant
Query response mode <newline>
mode is 0 or 1, and indicates that the automatic filter function is off or on,
respectively. Response data type is NR1.
Example :OUTP:FILT:AUTO 1
:OUTP2:FILT:AUTO?
:OUTPut:FILTer[:LPASs]:FREQuency
Sets the cutoff frequency of the output filter. This command setting is ignored if the
automatic filter function is enabled by the
:OUTPut:FILTer:AUTO command.
Syntax :OUTPut[c]:FILTer[:LPASs]:FREQuency frequency
:OUTPut[c]:FILTer[:LPASs]:FREQuency? [frequency]
Comments to this Manuals