4- 82 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:SENSe:DATA?
value for resistance measurement: See Table 2-9. This is available for the resistance
measurements set to the AUTO mode which is selected by the
:SENSe:RESistance:MODE command.
range=UP sets the next higher measurement range.
range=DOWN sets the next lower measurement range.
Query response range <newline>
range returns the present setting of the measurement range. If a parameter is
specified, range returns the value assigned to DEF, MIN, or MAX. Response data
type is NR3.
Example :SENS:CURR:RANG:UPP 1
:SENS2:CURR:DC:RANG:UPP?
:SENSe:DATA?
Returns the array data which contains all of the current measurement data, voltage
measurement data, resistance measurement data, source output setting data, status
data, or time data specified by the
:FORMat:ELEMents:SENSe command. The data
is not cleared until the :INITiate, :MEASure, or :READ command is executed.
Syntax :SENSe[c]:DATA? [offset[, size]]
Parameter offset Indicates the beginning of the data received. n|CURRent|STARt
(default). Parameter data type is NR1 or CPD.
offset=n specifies the n+1th data. n is an integer, 0 to maximum
(depends on the buffer state).
offset=CURR specifies the present data position.
offset=STAR specifies the top of the data buffer. Same as
offset=0.
size Number of data to be received. 1 to maximum (depends on the
buffer state). Parameter data type is NR1. If this parameter is
not specified, all data from offset is returned.
Query response response <newline>
response returns the array data specified by the :FORMat:ELEMents:SENSe
command. Response data type is NR3. See “Data Output Format” on page 1-12.