Agilent B2900 SCPI Command Reference, Edition 2 2- 21
Subsystem Command Summary
Setting Source/Measure Unit
:SENS[c]:WAIT:GAIN gain
:SENS[c]:WAIT:GAIN? [DEFault|MIN
imum|MAXimum]
Sets the gain value used for calculating
the measurement wait time for the
specified channel.
gain=MINimum|MAXimum|DEFault|
0
to 100
1
:SENS[c]:WAIT:OFFS offset
:SENS[c]:WAIT:OFFS? [DEFault|MIN
imum|MAXimum]
Sets the offset value used for calculating
the measurement wait time for the
specified channel.
gain=MINimum|MAXimum|DEFault|
0
to 1 seconds
0
:SENS[c]:WAIT[:STAT] mode
:SENS[c]:WAIT[:STAT]?
Enables or disables the measurement wait
time for the specified channel. The wait
time is defined as the time the
measurement channel cannot start
measurement after the start of a DC
output or the trailing edge of a pulse.
mode=0 or OFF disables the
measurement wait time. The wait time is
set to 0.
mode=1 or ON enables the measurement
wait time given by the following formula.
•:SENS[c]:WAIT:AUTO ON|1
condition:
wait time = gain initial wait time +
offset
•:SENS[c]:WAIT:AUTO OFF|0
condition:
wait time = offset
The initial wait time is automatically set
by the instrument and cannot be changed.
ON
Command Summary Reset setting