4- 10 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:CALCulate:LIMit:COMPliance:DIGital[:DATA]
:CALCulate[c]:FEED?
Parameter type Data type. MATH|RESistance|CURRent|VOLTage (default).
Parameter data type is CPD.
type=VOLT specifies the voltage measurement data.
type=CURR specifies the current measurement data.
type=RES specifies the resistance calculation data given by the following formula.
Resistance=Vmeas/Imeas
Where, Vmeas is the voltage measurement data, and Imeas is the current
measurement data.
type=MATH specifies the data given by a math expression. The math expression
must be specified before the :CALC:FEED MATH command is executed.
An existing math expression can be specified by the
:CALCulate:MATH[:EXPRession]:NAME command.
A new math expression can be defined by the
:CALCulate:MATH[:EXPRession]:NAME and
:CALCulate:MATH[:EXPRession][:DEFine] commands.
Query response type <newline>
type returns the present setting of data type, MATH, RES, CURR, or VOLT.
Response data type is CRD.
Example :CALC:FEED MATH
:CALC2:FEED?
:CALCulate:LIMit:COMPliance:DIGital[:DATA]
Defines the bit pattern used to indicate a failure of the compliance status check
specified by m. It must be entered in the format set by the
:FORMat:DIGital
command.
Syntax :CALCulate[c]:LIMit[m]:COMPliance:DIGital[:DATA] bit_pattern
:CALCulate[c]:LIMit[m]:COMPliance:DIGital[:DATA]?
Parameter bit_pattern 0 (default setting in decimal expression) to 16383. Parameter
data type is NR1 or NDN.