2- 18 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Command Summary
Setting Source/Measure Unit
:SENS[c]:RES:RANG:AUTO:ULIM rang
e
:SENS[c]:RES:RANG:AUTO:ULIM? [DE
Fault|MINimum|MAXimum]
:SENS[c]:CURR[:DC]:RANG:AUTO:ULI
M?
:SENS[c]:VOLT[:DC]:RANG:AUTO:ULI
M?
Specifies the upper limit for the automatic
measurement ranging operation, and sets
the maximum measurement range which
provides the best resolution to measure
the specified value. This is effective for
resistance measurements set to the AUTO
mode by the :SENS:RES:MODE
command.
range=MINimum|MAXimum|DEFault|
value (see
Table 2-9)
200 M
:SENS[c]:CURR[:DC]:RANG[:UPP] ra
nge
:SENS[c]:RES:RANG[:UPP] range
:SENS[c]:VOLT[:DC]:RANG[:UPP] ra
nge
:SENS[c]:CURR[:DC]:RANG[:UPP]? [
DEFault|MINimum|MAXimum]
:SENS[c]:RES:RANG[:UPP]? [DEFaul
t|MINimum|MAXimum]
:SENS[c]:VOLT[:DC]:RANG[:UPP]? [
DEFault|MINimum|MAXimum]
Specifies the expected measurement
value, and sets the measurement range
which provides the best resolution to
measure the specified value.
range=UP|DOWN|MINimum|MAXimu
m|DEFault| minimum to maximum
measurement value, in A, , or V. See
“Measurement Ranges” on page 2-27.
1.00E-04 A,
20 k, or 2
V
:SENS[c]:DATA? [offset[,size]]
Returns the array data which contains all
data for the element specified by the
:FORM:ELEM:SENS command.
offset=CURRent|STARt|0 to maximum
size=1 to maximum
STAR and
all
data
:SENS[c]:DATA:LAT?
Returns the latest data for the element
specified by the :FORM:ELEM:SENS
command.
:SENS[c]:FUNC:OFF fctn[,fctn[,fc
tn]]
:SENS[c]:FUNC:OFF?
Disables the specified measurement
functions.
fctn=“CURRent[:DC]”|“VOLTage[:DC]”
|“RESistance”
“RES”
Command Summary Reset setting