Agilent B2900 SCPI Command Reference, Edition 2 2- 13
Subsystem Command Summary
Setting Source/Measure Unit
[:SOUR[c]]:SWE:RANG mode
[:SOUR[c]]:SWE:RANG?
Selects the output ranging mode of the
sweep output for the specified channel.
mode=BEST|FIXed|AUTO
BEST
[:SOUR[c]]:SWE:SPAC mode
[:SOUR[c]]:SWE:SPAC?
Selects the scale of the sweep output for
the specified channel. See
Figure 2-3.
mode=LOGarithmic|LINear
LIN
[:SOUR[c]]:SWE:STA mode
[:SOUR[c]]:SWE:STA?
Sets the sweep mode for the specified
channel.
mode=SINGle|DOUBle
SING
[:SOUR[c]]:TOUT:SIGN output{,out
put}
[:SOUR[c]]:TOUT:SIGN output{,out
put}
[:SOUR[c]]:TOUT:SIGN output{,out
put}
[:SOUR[c]]:TOUT:SIGN?
[:SOUR[c]]:TOUT:SIGN?
Selects the trigger output for the status
change between the trigger layer and the
transient device action.
output=INT1|INT2|LAN|EXT1|EXT2|
EXT3|EXT4|EXT5|EXT6|EXT7|EXT8|
EXT9|EXT10|EXT11|EXT12|EXT13|
EXT14
EXT1
[:SOUR[c]]:TOUT[:STAT] mode
[:SOUR[c]]:TOUT[:STAT] mode
[:SOUR[c]]:TOUT[:STAT] mode
[:SOUR[c]]:TOUT[:STAT]?
[:SOUR[c]]:TOUT[:STAT]?
Enables or disables the trigger output for
the status change between the trigger
layer and the transient device action.
mode=1|ON|0|OFF
OFF
[:SOUR[c]]:WAIT:AUTO mode
[:SOUR[c]]:WAIT:AUTO?
Enables or disables the initial wait time
used for calculating the source wait time
for the specified channel. See
[:SOUR[c]]:WAIT[:STAT].
mode=1|ON|0|OFF
ON
Command Summary Reset setting