100 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
4 BW/Avg
Avg/VBW Type
Displays the functions that enable you to automatically or manually choose one of the
following averaging scales: log-power averaging (video), power averaging (RMS), or voltage
averaging.
When you select log-power averaging, the measurement results are the average of the
signal level in logarithmic units (decibels). When you select power averaging (RMS), all
measured results are converted into power units before averaging and filtering operations,
and converted back to decibels for displaying. Remember: there can be significant
differences between the average of the log of power and the log of the average power.
The following are the averaging processes within the N8201A, all of which are affected by
this setting:
Trace averaging (see BW/Avg) averages signal amplitudes on a trace-to-trace basis.
Average detector (see Detector, Average) averages signal amplitudes during the time or
frequency interval represented by a particular measurement point.
Noise Marker (see Marker Noise) averages signal amplitudes across measurement points
to reduce variations for noisy signals.
VBW filtering adds video filtering which is a form of averaging of the video signal.
When Auto is selected, averaging for the current instrument measurement settings is
chosen
Auto selects Power (RMS) Averaging if Marker Noise is on, Band/Intvl Power is on, or
Detector is set to Man and Average. It selects Voltage Averaging if Amplitude, Scale Type is
set to Lin or Detector is set to Quasi Peak or EMI Average. For other conditions, Auto
selects Log-Power Average.
When manual is selected, the type is shown on the left side of the display with a #. When
auto is selected, the N8201A chooses the type of averaging. When one of the average types
is selected manually, the N8201A uses that type regardless of other N8201A settings, and
sets Avg/VBW Type to Man.
Key Path: BW/Avg or Auto Couple
Saved State: Saved in Instrument State
Factory Preset: Auto (Log-power)
Remote Command:
[:SENSe]:AVERage:TYPE RMS|LOG|SCALar
[:SENSe]:AVERage:TYPE?
[:SENSe]:AVERage:TYPE:AUTO OFF|ON|0|1