562 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
27 Using the STATus System
Key Type: There is no equivalent front-panel key.
Operation Negative Transition
:STATus:OPERation:NTRansition <integer>
:STATus:OPERation:NTRansition?
Determines which bits in the Operation Condition register will set the corresponding bit in
the Operation Event register when the condition register bit has a negative transition (1 to
0). The variable <number> is the sum of the decimal values of the bits that you want to
enable.
Key Type: There is no equivalent front-panel key.
Factory Preset: 0
Range: 0 to 32767
Operation Positive Transition
:STATus:OPERation:PTRansition <integer>
:STATus:OPERation:PTRansition?
Determines which bits in the Operation Condition register will set the corresponding bit in
the Operation Event register when the condition register bit has a positive transition (0 to
1). The variable <number> is the sum of the decimal values of the bits that you want to
enable.
Key Type: There is no equivalent front-panel key.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767
Preset the Status Byte
:STATus:PRESet
Sets bits in most of the enable and transition registers to their default state. It presets all
the Transition Filters, Enable Registers, and the Error/Event Queue Enable. It has no effect
on Event Registers, Error/Event QUEue, IEEE 488.2 ESE, and SRE Registers as described in
IEEE Standard 488.2-1992, IEEE Standard Codes, Formats, Protocols and Common
Commands for Use with ANSI/IEEE Std 488.1-1987. New York, NY, 1992.
Key Type: There is no equivalent front-panel key.